| Phenomenon | Cause | Solution | 
| 1. No power indication | 1.The power cord is not plugged in or disconnected 2.Open fuse | 1.Insert the power plug, check the power cord 2. Replace the fuse | 
| 2. The temperature in the box does not rise | 1. The set temperature is too low 2. The temperature controller is damaged. 3.The electric heater is damaged | 1. Adjust the set temperature 2. Replace the temperature controller 3. Replace the electric heater 
 | 
| 3. The PV of the meter displays "0000" or the numbers are string up and down. | The temperature sensor is damaged. | Replace the temperature sensor. | 
| 4. Abnormal over temperature alarm | 1. The set temperature is too low 2. The temperature controller is damaged | 1. Adjust the set temperature 2.Replace the temperature controller 
 | 
| 5. Humidity does not go up | Over-temperature setting is too low | Set over-temperature value correctly, between 55-60 | 
| The fan does not rotate | Check and confirm that the fan resistance is too large or the capacitor is damaged, and replace the fan or capacitor of the same type | |
| There is a problem with the sensor | Replace the sensor of the same type | |
| The water level of the water tank is not enough; the position of the water cup is high. | Fill up with water; Adjust the position of the water cup 
 | |
| There is no water in the humidification shallow water tank, and the water level alarm will act. | Check the waterway to make the waterway unblocked. The float action should have on-off action, without replacing the float. | |
| Wet bulb sensor is deviated | Replace sensor | |
| 6. The evaporator is frosted or the inner liner is frosted. | Open the door frequently when it is at low temperature. | Stop cooling and dry it at 80 degrees, which may reduce the number of door openings. | 
| The lead hole is not tightly sealed | Cover with rubber stopper during low temperature 
 |